In a 3:1 scaled model of the human aorta models of the Omniscience, Bjo¨rk-Shiley Convexo-Concave, Bjo¨rk-Shiley Monostrut, Medtronic-Hall, Duromedics (Hemex) and the Saint Jude Medical heart valve prostheses are studied in steady flow representing the systolic peak flow phase. Detailed flow visualization experiments show flow separations at all inner ring surfaces as well as at most of the occluders. The resulting stagnation areas increase the risk of thrombus accumulation. Flow separations also stimulate vortex formation and turbulent mixing at the downstream jet boundaries and thus may intensify blood damage by turbulent shear stresses. The different influences of struts and occluder guides on the flow around the occluders are discussed. The effects of the individual valve components on the flow fields are analyzed and correlated with the resulting pressure losses.
Skip Nav Destination
Article navigation
November 1988
Research Papers
Model Studies at Mechanical Aortic Heart Valve Prostheses—Part I: Steady-State Flow Fields and Pressure Loss Coefficients
Martin Knoch,
Martin Knoch
Helmholtz-Institute for Biomedical Engineering, at the Aachen University of Technology, D-5100 Aachen, Federal Republic of Germany
Search for other works by this author on:
Helmut Reul,
Helmut Reul
Helmholtz-Institute for Biomedical Engineering, at the Aachen University of Technology, D-5100 Aachen, Federal Republic of Germany
Search for other works by this author on:
Ralf Kro¨ger,
Ralf Kro¨ger
Helmholtz-Institute for Biomedical Engineering, at the Aachen University of Technology, D-5100 Aachen, Federal Republic of Germany
Search for other works by this author on:
Gu¨nter Rau
Gu¨nter Rau
Helmholtz-Institute for Biomedical Engineering, at the Aachen University of Technology, D-5100 Aachen, Federal Republic of Germany
Search for other works by this author on:
Martin Knoch
Helmholtz-Institute for Biomedical Engineering, at the Aachen University of Technology, D-5100 Aachen, Federal Republic of Germany
Helmut Reul
Helmholtz-Institute for Biomedical Engineering, at the Aachen University of Technology, D-5100 Aachen, Federal Republic of Germany
Ralf Kro¨ger
Helmholtz-Institute for Biomedical Engineering, at the Aachen University of Technology, D-5100 Aachen, Federal Republic of Germany
Gu¨nter Rau
Helmholtz-Institute for Biomedical Engineering, at the Aachen University of Technology, D-5100 Aachen, Federal Republic of Germany
J Biomech Eng. Nov 1988, 110(4): 334-343 (10 pages)
Published Online: November 1, 1988
Article history
Received:
December 21, 1987
Revised:
July 20, 1988
Online:
June 12, 2009
Citation
Knoch, M., Reul, H., Kro¨ger, R., and Rau, G. (November 1, 1988). "Model Studies at Mechanical Aortic Heart Valve Prostheses—Part I: Steady-State Flow Fields and Pressure Loss Coefficients." ASME. J Biomech Eng. November 1988; 110(4): 334–343. https://doi.org/10.1115/1.3108450
Download citation file:
Get Email Alerts
Related Articles
Reduction of Procoagulant Potential of b-Datum Leakage Jet Flow in Bileaflet Mechanical Heart Valves via Application of Vortex Generator Arrays
J Biomech Eng (July,2010)
Influence of Cardiac Flow Rate on Turbulent Shear Stress from a Prosthetic Heart Valve
J Biomech Eng (May,1988)
Two-Component Laser Velocimeter Measurements Downstream of Heart Valve Prostheses in Pulsatile Flow
J Biomech Eng (February,1986)
Vortex Shedding as a Mechanism for Free Emboli Formation in Mechanical Heart Valves
J Biomech Eng (April,2000)
Related Proceedings Papers
Related Chapters
Introduction
Mechanical Blood Trauma in Circulatory-Assist Devices
Concluding remarks
Mechanical Blood Trauma in Circulatory-Assist Devices
Dynamic Behavior of Pumping Systems
Pipeline Pumping and Compression Systems: A Practical Approach